PlainRecalls
FDA Devices Moderate Class II Terminated

Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: "Querying and retrieving patient history information and/or previo

Reported: October 8, 2014 Initiated: September 2, 2014 #Z-0004-2015

Product Description

Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: "Querying and retrieving patient history information and/or previous diagnosis and images from other modalities; "X-ray examinations of the urogenital area "Ultrasound examinations "Endourological interventions "Percutaneous interventions "Laparoscopy "Application of fistula "Simple procedures "Extracorporeal shock wave lithotripsy "Uroflow/urodynamics "Pediatric radiological and therapeutic applications

Reason for Recall

Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

Details

Units Affected
2
Distribution
Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: "Querying and retrieving patient history information and/or previous diagnosis and images from other modalities; "X-ray examinations of the urogenital area "Ultrasound examinations "Endourological interventions "Percutaneous interventions "Laparoscopy "Application of fistula "Simple procedures "Extracorporeal shock wave lithotripsy "Uroflow/urodynamics "Pediatric radiological and therapeutic applications. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 2.
Why was this product recalled?
Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
Which agency issued this recall?
This recall was issued by the FDA Devices on October 8, 2014. Severity: Moderate. Recall number: Z-0004-2015.