Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Reported: October 8, 2025 Initiated: August 29, 2025 #Z-0016-2026
Product Description
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Reason for Recall
Potential for device to overheat during operation f the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
Details
- Recalling Firm
- Getinge Disinfection Ab
- Units Affected
- 877 units
- Distribution
- Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Location
- Vaxjo, N/A
Frequently Asked Questions
What product was recalled? ▼
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.. Recalled by Getinge Disinfection Ab. Units affected: 877 units.
Why was this product recalled? ▼
Potential for device to overheat during operation f the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on October 8, 2025. Severity: Moderate. Recall number: Z-0016-2026.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11