OPTETRAK LOGIC Polyethylene Tibial Inserts, labeled as the following: a. OPTETRAK Logic PS Tibial Inserts, Product Line 02-012-35-XXXX; b. OPTETRAK Logic PSC Tibial Inserts, Product Line 02-012-44-XXXX; c. OPTETRAK Logic CR Tibial Inserts, Product Line 02-012-47-XXXX; d. OPTETRAK Logic CR Slope + Tibial Inserts, Product Line 02-012-48-XXXX; e. OPTETRAK Logic CR Slope ++ Tibial Inserts, Product Line 02-012-49-XXXX; f. OPTETRAK Logic CRC Tibial Inserts, Product Line 02
Product Description
OPTETRAK LOGIC Polyethylene Tibial Inserts, labeled as the following: a. OPTETRAK Logic PS Tibial Inserts, Product Line 02-012-35-XXXX; b. OPTETRAK Logic PSC Tibial Inserts, Product Line 02-012-44-XXXX; c. OPTETRAK Logic CR Tibial Inserts, Product Line 02-012-47-XXXX; d. OPTETRAK Logic CR Slope + Tibial Inserts, Product Line 02-012-48-XXXX; e. OPTETRAK Logic CR Slope ++ Tibial Inserts, Product Line 02-012-49-XXXX; f. OPTETRAK Logic CRC Tibial Inserts, Product Line 02-012-51-XXXX; g. OPTETRAK Logic CC Tibial Inserts, Product Line 02-012-65-XXXX; Affected 510(k) numbers include: K033883, K132161, K093360, K110547, K111400, K121307, K123342, K150890. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.
Reason for Recall
Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Details
- Recalling Firm
- Exactech, Inc.
- Units Affected
- 109,787 devices
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, ECUADOR, FRANCE, FRENCH POLYNESIA, GERMANY, GHANA, GREAT BRITAIN, GREECE, GUATEMALA, INDIA, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, ROMANIA, SINGAPORE, SPAIN, SWITZERLAND, SYRIAN ARAB REPUBLIC, THAILAND, TUNISIA, TURKEY, VENEZUELA .
- Location
- Gainesville, FL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11