PlainRecalls
FDA Devices Moderate Class II Ongoing

FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

Reported: October 16, 2024 Initiated: August 26, 2024 #Z-0022-2025

Product Description

FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

Reason for Recall

Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.

Details

Recalling Firm
Qualigen Inc
Units Affected
60
Distribution
US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland
Location
Carlsbad, CA

Frequently Asked Questions

What product was recalled?
FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer. Recalled by Qualigen Inc. Units affected: 60.
Why was this product recalled?
Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 16, 2024. Severity: Moderate. Recall number: Z-0022-2025.