Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use
Product Description
Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use
Reason for Recall
IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.
Details
- Recalling Firm
- Percussionaire Corporation
- Units Affected
- 59160
- Distribution
- Worldwide distribution. US nationwide, Russia, Japan, Switzerland, Canada, Qatar, Chile, South Africa, United Kingdom, Israel, Turkey, and UAE
- Location
- Sandpoint, ID
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11