Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1

Reported: October 16, 2024 Initiated: April 19, 2023 #Z-0035-2025

Product Description

Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1

Reason for Recall

During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

Details

Recalling Firm
Philips North America
Units Affected
591 devices
Distribution
Worldwide - US Nationwide distribution.
Location
Cambridge, MA

Frequently Asked Questions

What product was recalled?
Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1. Recalled by Philips North America. Units affected: 591 devices.
Why was this product recalled?
During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 16, 2024. Severity: Moderate. Recall number: Z-0035-2025.