GE Heathcare Giraffe OmniBed; neo natal incubator

Reported: October 23, 2024 Initiated: September 3, 2024 #Z-0047-2025

Product Description

GE Heathcare Giraffe OmniBed; neo natal incubator

Reason for Recall

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

Details

Recalling Firm
DATEX--OHMEDA, INC.
Distribution
Worldwide distribution.
Location
Wauwatosa, WI

Frequently Asked Questions

What product was recalled?
GE Heathcare Giraffe OmniBed; neo natal incubator. Recalled by DATEX--OHMEDA, INC..
Why was this product recalled?
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 23, 2024. Severity: Critical. Recall number: Z-0047-2025.