Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage¿ Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage¿ TOX Drug Screen package insert, p
Product Description
Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage¿ Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage¿ TOX Drug Screen package insert, provide this capability. The results obtained for the controls are to be compared with the assigned values given on the Expected Values Card, accompanying the package insert, to determine if the procedure is within control limits. Alere Triage¿ TOX Drug Screen Control 1 or Alere Triage¿ TOX Drug Screen Control 2 consist of human urine matrix containing parent compounds and/or major metabolites/analogues of drugs of abuse. 510(k)/IDE/PMA number: K060788 Type of Packaging: 5 vials per level in a labeled kit box. Indications for use: The Alere Triage¿ TOX Drug Screen Controls are to be used with the Alere Triage¿ TOX Drug Screen tests and Alere Triage¿ Meters to assist the laboratory in monitoring test performance.
Reason for Recall
Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.
Details
- Recalling Firm
- Alere San Diego, Inc.
- Units Affected
- 38
- Distribution
- Worldwide Distribution - USA (nationwide) and countries of: South Africa and Canada.
- Location
- San Diego, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11