Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage¿ Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage¿ TOX Drug Screen package insert, p

Reported: October 30, 2013 Initiated: September 23, 2013 #Z-0053-2014

Product Description

Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage¿ Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage¿ TOX Drug Screen package insert, provide this capability. The results obtained for the controls are to be compared with the assigned values given on the Expected Values Card, accompanying the package insert, to determine if the procedure is within control limits. Alere Triage¿ TOX Drug Screen Control 1 or Alere Triage¿ TOX Drug Screen Control 2 consist of human urine matrix containing parent compounds and/or major metabolites/analogues of drugs of abuse. 510(k)/IDE/PMA number: K060788 Type of Packaging: 5 vials per level in a labeled kit box. Indications for use: The Alere Triage¿ TOX Drug Screen Controls are to be used with the Alere Triage¿ TOX Drug Screen tests and Alere Triage¿ Meters to assist the laboratory in monitoring test performance.

Reason for Recall

Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.

Details

Recalling Firm
Alere San Diego, Inc.
Units Affected
38
Distribution
Worldwide Distribution - USA (nationwide) and countries of: South Africa and Canada.
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage¿ Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage¿ TOX Drug Screen package insert, provide this capability. The results obtained for the controls are to be compared with the assigned values given on the Expected Values Card, accompanying the package insert, to determine if the procedure is within control limits. Alere Triage¿ TOX Drug Screen Control 1 or Alere Triage¿ TOX Drug Screen Control 2 consist of human urine matrix containing parent compounds and/or major metabolites/analogues of drugs of abuse. 510(k)/IDE/PMA number: K060788 Type of Packaging: 5 vials per level in a labeled kit box. Indications for use: The Alere Triage¿ TOX Drug Screen Controls are to be used with the Alere Triage¿ TOX Drug Screen tests and Alere Triage¿ Meters to assist the laboratory in monitoring test performance.. Recalled by Alere San Diego, Inc.. Units affected: 38.
Why was this product recalled?
Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 30, 2013. Severity: Low. Recall number: Z-0053-2014.