Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-imaging data . Enterprise Imaging XERO Viewer 8.0.0, 8.0.1, 8.1, 8.1.1, 8.1.2, 8.1.3,8.1.4, 8.2.0.000, 8.2.1.000, 8.2.2.000, 8.3.0.000, 8.3.1.000, 8.3.2.000, 8.4.0.000, 8.4.1.000 (including all SU s and HF s)
Product Description
Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-imaging data . Enterprise Imaging XERO Viewer 8.0.0, 8.0.1, 8.1, 8.1.1, 8.1.2, 8.1.3,8.1.4, 8.2.0.000, 8.2.1.000, 8.2.2.000, 8.3.0.000, 8.3.1.000, 8.3.2.000, 8.4.0.000, 8.4.1.000 (including all SU s and HF s)
Reason for Recall
Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.
Details
- Recalling Firm
- Agfa Healthcare NV
- Units Affected
- 801 units
- Distribution
- US Nationwide distribution. Foreign: Argentina, Australia, Belgium, Chile, Canada, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Georgia, Greece, Hong Kong, Iceland, Ireland, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Nicaragua, Norway, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Tanzania, Tunisia, United Arab Emirates, United Kingdom
- Location
- Mortsel
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11