PROPONENT SR SL (Model L200)

Reason for Recall

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Details

Units Affected
1380 units
Distribution
Worldwide
Location
Saint Paul, MN

Frequently Asked Questions

What product was recalled?
PROPONENT SR SL (Model L200). Recalled by Boston Scientific Corporation. Units affected: 1380 units.
Why was this product recalled?
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Which agency issued this recall?
This recall was issued by the FDA Devices on October 22, 2025. Severity: Critical. Recall number: Z-0085-2026.