Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A

Reported: October 26, 2022 Initiated: August 24, 2022 #Z-0089-2023

Product Description

Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A

Reason for Recall

Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.

Details

Units Affected
1592 units
Distribution
Nationwide Foreign: Canada
Location
Newark, DE

Frequently Asked Questions

What product was recalled?
Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 1592 units.
Why was this product recalled?
Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 26, 2022. Severity: Moderate. Recall number: Z-0089-2023.