Symbia Intevo 6, SPECT/CT System, Emission Computed Tomography System To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. CT: The CT com
Product Description
Symbia Intevo 6, SPECT/CT System, Emission Computed Tomography System To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. CT: The CT component is intended to produce cross sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial place taken at different angles or spiral planes taken at different angles.SPECT+CT: Perform CT scans and nuclear imaging studies within the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
Reason for Recall
The Broad Quantification option of the Symbia product software version VB20A may not allow modification of two data input values. The failure occurs when the system with the Broad Quantification option is calibrated with phantoms that do not have a volume of 6500ml and/or when residual dose needs to be entered.
Details
- Recalling Firm
- Siemens Medical Solutions USA, Inc.
- Units Affected
- 2 units
- Distribution
- Worldwide Distribution to Malaysia, France, Japan, Reunion Island
- Location
- Hoffman Estates, IL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11