PlainRecalls
FDA Devices Moderate Class II Terminated

AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz -Fresenius Part Number: 24-1500-1 AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz- Fresenius Part Number: 24-2000-0

Reported: October 21, 2020 Initiated: September 1, 2020 #Z-0098-2021

Product Description

AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz -Fresenius Part Number: 24-1500-1 AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz- Fresenius Part Number: 24-2000-0

Reason for Recall

Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.

Details

Units Affected
15 units
Distribution
US distribution to DE, NC, OK, PA, and TX.
Location
Waltham, MA

Frequently Asked Questions

What product was recalled?
AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz -Fresenius Part Number: 24-1500-1 AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz- Fresenius Part Number: 24-2000-0. Recalled by Fresenius Medical Care Holdings, Inc.. Units affected: 15 units.
Why was this product recalled?
Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 21, 2020. Severity: Moderate. Recall number: Z-0098-2021.