Product Description

Access Thyroglobulin Calibrators

Reason for Recall

It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.

Details

Recalling Firm
Beckman Coulter, Inc.
Units Affected
5980 units
Distribution
worldwide except US, Canada, and China
Location
Chaska, MN

Frequently Asked Questions

What product was recalled?
Access Thyroglobulin Calibrators. Recalled by Beckman Coulter, Inc.. Units affected: 5980 units.
Why was this product recalled?
It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.