FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic
Reported: October 15, 2025 Initiated: September 3, 2025 #Z-0107-2026
Product Description
FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic
Reason for Recall
The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
Details
- Recalling Firm
- Inpeco S.A.
- Units Affected
- 1 unit
- Distribution
- US distribution to CA & NY.
- Location
- Lugano
Frequently Asked Questions
What product was recalled? ▼
FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic. Recalled by Inpeco S.A.. Units affected: 1 unit.
Why was this product recalled? ▼
The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on October 15, 2025. Severity: Moderate. Recall number: Z-0107-2026.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11