PlainRecalls
FDA Devices Moderate Class II Ongoing

Aestiva 7900, Model Numbers: a) 1006-9300-000 b) 1006-9305-000 c) 1006-9118-000 d) 1006-9112-000 e) 1006-9113-000 f) 1006-9020-000 g) 1006-9033-000 h) 1006-9117-000 i) 1006-9119-000 j) 1006-9030-000 k) 1006-9032-000 l) 1006-9022-000 m) To be provided n) 1006-9120-000 o) 1006-9027-000 p) 1006-9003-000 q) 1006-9025-000

Reported: December 25, 2019 Initiated: July 19, 2019 #Z-0115-2020

Product Description

Aestiva 7900, Model Numbers: a) 1006-9300-000 b) 1006-9305-000 c) 1006-9118-000 d) 1006-9112-000 e) 1006-9113-000 f) 1006-9020-000 g) 1006-9033-000 h) 1006-9117-000 i) 1006-9119-000 j) 1006-9030-000 k) 1006-9032-000 l) 1006-9022-000 m) To be provided n) 1006-9120-000 o) 1006-9027-000 p) 1006-9003-000 q) 1006-9025-000

Reason for Recall

Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.

Details

Recalling Firm
GE Healthcare, LLC
Units Affected
7168 devices
Distribution
Worldwide
Location
Waukesha, WI

Frequently Asked Questions

What product was recalled?
Aestiva 7900, Model Numbers: a) 1006-9300-000 b) 1006-9305-000 c) 1006-9118-000 d) 1006-9112-000 e) 1006-9113-000 f) 1006-9020-000 g) 1006-9033-000 h) 1006-9117-000 i) 1006-9119-000 j) 1006-9030-000 k) 1006-9032-000 l) 1006-9022-000 m) To be provided n) 1006-9120-000 o) 1006-9027-000 p) 1006-9003-000 q) 1006-9025-000. Recalled by GE Healthcare, LLC. Units affected: 7168 devices.
Why was this product recalled?
Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 25, 2019. Severity: Moderate. Recall number: Z-0115-2020.