PlainRecalls
FDA Devices Moderate Class II Ongoing

Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR RED Model/Catalog Number: 4329RD Product Description: ROLLING WALKER MOBILITY AID

Reported: October 23, 2024 Initiated: August 26, 2024 #Z-0131-2025

Product Description

Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR RED Model/Catalog Number: 4329RD Product Description: ROLLING WALKER MOBILITY AID

Reason for Recall

Firm identified a potential risk for a fall hazard. When the end user is readjusting from the slouched seated position to an upright seated position, there is the potential for the movement to cause the backrest to lift out of the backrest holder and disengage from the unit.

Details

Recalling Firm
Nova Ortho-Med Inc
Units Affected
1585
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL AZ, CA, CO, CT, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WI and WY. The country of Canada.
Location
Carson, CA

Frequently Asked Questions

What product was recalled?
Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR RED Model/Catalog Number: 4329RD Product Description: ROLLING WALKER MOBILITY AID. Recalled by Nova Ortho-Med Inc. Units affected: 1585.
Why was this product recalled?
Firm identified a potential risk for a fall hazard. When the end user is readjusting from the slouched seated position to an upright seated position, there is the potential for the movement to cause the backrest to lift out of the backrest holder and disengage from the unit.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 23, 2024. Severity: Moderate. Recall number: Z-0131-2025.