PlainRecalls
FDA Devices Moderate Class II Terminated

Transfer/Gait Belt, Plastic, Wipeable, packaged in plastic bag, 36 plastic bags per box.

Reported: November 29, 2017 Initiated: August 10, 2017 #Z-0164-2018

Product Description

Transfer/Gait Belt, Plastic, Wipeable, packaged in plastic bag, 36 plastic bags per box.

Reason for Recall

The locking mechanism on the belt may fail to secure/hold the strap when in locked position potentially allowing the belt strap to slip through the buckle.

Details

Recalling Firm
Medline Industries Inc
Units Affected
1,428 units total
Distribution
USA (nationwide) Distribution to the states of : AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NV, NY, OH , OK, OR, PA, RI, SC, TN, TX, VA, VT, WA, WI and WV.
Location
Northfield, IL

Frequently Asked Questions

What product was recalled?
Transfer/Gait Belt, Plastic, Wipeable, packaged in plastic bag, 36 plastic bags per box.. Recalled by Medline Industries Inc. Units affected: 1,428 units total.
Why was this product recalled?
The locking mechanism on the belt may fail to secure/hold the strap when in locked position potentially allowing the belt strap to slip through the buckle.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 29, 2017. Severity: Moderate. Recall number: Z-0164-2018.