PlainRecalls
FDA Devices Moderate Class II Terminated

6" (15 cm) Appx 0.26 ml, Smallbore Ext Set w/Clamp, NanoClave T-Connector, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case; UDI:(01)10887709051024(17)250101(30)50(10)4603240;(01)10887709051024(17)250201(30)50(10)4714349; (01)10887709051024(17)250201(30)50(10)4733540; (01)10887709051024(17)250201(30)50(10)4724812; (01)10887709051024(17)241201(30)50(10)4515226; (01)10887709051024(17)241201(30)50(10)4537113; (01)10887709051024(17)250101(30)50(10)4557474; (01)10887

Reported: October 28, 2020 Initiated: July 6, 2020 #Z-0183-2021

Product Description

6" (15 cm) Appx 0.26 ml, Smallbore Ext Set w/Clamp, NanoClave T-Connector, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case; UDI:(01)10887709051024(17)250101(30)50(10)4603240;(01)10887709051024(17)250201(30)50(10)4714349; (01)10887709051024(17)250201(30)50(10)4733540; (01)10887709051024(17)250201(30)50(10)4724812; (01)10887709051024(17)241201(30)50(10)4515226; (01)10887709051024(17)241201(30)50(10)4537113; (01)10887709051024(17)250101(30)50(10)4557474; (01)10887709051024(17)250201(30)50(10)4718500; (01)10887709052793(17)250101(30)50(10)4557476; (01)10887709052793(17)250101(30)50(10)4590957; The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Reason for Recall

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Details

Recalling Firm
ICU Medical, Inc.
Units Affected
Total of all products (Listed #1 thru 101) = 304735 units
Distribution
Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Location
San Clemente, CA

Frequently Asked Questions

What product was recalled?
6" (15 cm) Appx 0.26 ml, Smallbore Ext Set w/Clamp, NanoClave T-Connector, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case; UDI:(01)10887709051024(17)250101(30)50(10)4603240;(01)10887709051024(17)250201(30)50(10)4714349; (01)10887709051024(17)250201(30)50(10)4733540; (01)10887709051024(17)250201(30)50(10)4724812; (01)10887709051024(17)241201(30)50(10)4515226; (01)10887709051024(17)241201(30)50(10)4537113; (01)10887709051024(17)250101(30)50(10)4557474; (01)10887709051024(17)250201(30)50(10)4718500; (01)10887709052793(17)250101(30)50(10)4557476; (01)10887709052793(17)250101(30)50(10)4590957; The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).. Recalled by ICU Medical, Inc.. Units affected: Total of all products (Listed #1 thru 101) = 304735 units.
Why was this product recalled?
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 28, 2020. Severity: Moderate. Recall number: Z-0183-2021.