FDA Devices Critical Class I Ongoing

BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT160

Reported: November 15, 2023 Initiated: October 12, 2023 #Z-0189-2024

Product Description

BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT160

Reason for Recall

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Details

Units Affected
15942 devices
Distribution
US Nationwide distribution
Location
Center Valley, PA

Frequently Asked Questions

What product was recalled?
BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT160. Recalled by Olympus Corporation of the Americas. Units affected: 15942 devices.
Why was this product recalled?
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 15, 2023. Severity: Critical. Recall number: Z-0189-2024.