PlainRecalls
FDA Devices Moderate Class II Terminated

BD Vacutainer¿ Push Button Blood Collection Set with Pre-Attached Holder Catalog number 368356 - Product Usage: This product is designed to provide customers an increased level of safety during IV blood collection, convenience and to improve OSHA single-use holder compliance. Figure 1 shows a BD Vacutainer¿ Push Button Blood Collection Set with the needle protector removed prior to needle retraction.

Reported: October 30, 2019 Initiated: July 31, 2019 #Z-0193-2020

Product Description

BD Vacutainer¿ Push Button Blood Collection Set with Pre-Attached Holder Catalog number 368356 - Product Usage: This product is designed to provide customers an increased level of safety during IV blood collection, convenience and to improve OSHA single-use holder compliance. Figure 1 shows a BD Vacutainer¿ Push Button Blood Collection Set with the needle protector removed prior to needle retraction.

Reason for Recall

Separation of front and rear barrels upon activation of the safety feature that retracts the needle

Details

Units Affected
1735600 units
Distribution
Worldwide distribution - US Nationwide distribution including the states of CT, FL, MO, TN, OH, TX, GA, NC, SC, IN, IL, PA, CO and countries of Trinidad and Tobago, Hong Kong.
Location
Franklin Lakes, NJ

Frequently Asked Questions

What product was recalled?
BD Vacutainer¿ Push Button Blood Collection Set with Pre-Attached Holder Catalog number 368356 - Product Usage: This product is designed to provide customers an increased level of safety during IV blood collection, convenience and to improve OSHA single-use holder compliance. Figure 1 shows a BD Vacutainer¿ Push Button Blood Collection Set with the needle protector removed prior to needle retraction.. Recalled by Becton Dickinson & Company. Units affected: 1735600 units.
Why was this product recalled?
Separation of front and rear barrels upon activation of the safety feature that retracts the needle
Which agency issued this recall?
This recall was issued by the FDA Devices on October 30, 2019. Severity: Moderate. Recall number: Z-0193-2020.