PlainRecalls
FDA Devices Moderate Class II Ongoing

VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code: 8568040 . VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of ACET, CRBM, DGXN, PHBR, and PHYT

Reported: November 16, 2022 Initiated: October 13, 2022 #Z-0220-2023

Product Description

VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code: 8568040 . VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of ACET, CRBM, DGXN, PHBR, and PHYT

Reason for Recall

Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.

Details

Units Affected
17,396 Units
Distribution
Worldwide Distribution: US (Nationwide) and Countries (Foreign) including: Australia Bermuda DV04 Bermuda, HM 13 Bermuda, HM08 Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden United Kingdom
Location
Raritan, NJ

Frequently Asked Questions

What product was recalled?
VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code: 8568040 . VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of ACET, CRBM, DGXN, PHBR, and PHYT. Recalled by Ortho-Clinical Diagnostics, Inc.. Units affected: 17,396 Units.
Why was this product recalled?
Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 16, 2022. Severity: Moderate. Recall number: Z-0220-2023.