FDA Devices Moderate Class II Completed

A1CNow+ Multi-Test HbA1c System consists of (1) a semi-disposable plastic-encased device (the monitor), (2) a plastic cartridge enclosing dry reagent strips, and (3) a sample dilution kit for: collecting the blood sample, mixing the sample with the required pre-treatment solution, and delivering the sample to the cartridge.

Reported: October 30, 2024 Initiated: September 18, 2024 #Z-0223-2025

Product Description

A1CNow+ Multi-Test HbA1c System consists of (1) a semi-disposable plastic-encased device (the monitor), (2) a plastic cartridge enclosing dry reagent strips, and (3) a sample dilution kit for: collecting the blood sample, mixing the sample with the required pre-treatment solution, and delivering the sample to the cartridge.

Reason for Recall

The reason for this recall is PTS Diagnostics, Inc. received complaints of non-matching cartridge and analyzer codes within the product test system box. Use of cartridges and analyzers with non-matching codes may result in inaccurate test results as components within each lot number are meant to be used together, and not mixed between lots. Therefore, the test results may possibly lead to Hemoglobin A1C positive or negative bias results. While the product is for monitoring only, over recovery with a higher than accurate value can be an indication for increased therapy; such as physical activity, weight reduction, meal management, pharmaceutical therapy.

Details

Units Affected
2,452 units
Distribution
Worldwide distribution - US Nationwide and the countries of Japan, South Africa.
Location
Whitestown, IN

Frequently Asked Questions

What product was recalled?
A1CNow+ Multi-Test HbA1c System consists of (1) a semi-disposable plastic-encased device (the monitor), (2) a plastic cartridge enclosing dry reagent strips, and (3) a sample dilution kit for: collecting the blood sample, mixing the sample with the required pre-treatment solution, and delivering the sample to the cartridge.. Recalled by Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc.. Units affected: 2,452 units.
Why was this product recalled?
The reason for this recall is PTS Diagnostics, Inc. received complaints of non-matching cartridge and analyzer codes within the product test system box. Use of cartridges and analyzers with non-matching codes may result in inaccurate test results as components within each lot number are meant to be used together, and not mixed between lots. Therefore, the test results may possibly lead to Hemoglobin A1C positive or negative bias results. While the product is for monitoring only, over recovery with a higher than accurate value can be an indication for increased therapy; such as physical activity, weight reduction, meal management, pharmaceutical therapy.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 30, 2024. Severity: Moderate. Recall number: Z-0223-2025.