Synchro2, 0.014in x 300cm STANDARD Guidewire with Hydrophilic Coating UPN Product No. M00326510, REF 2651, STERILE:EO - Product Usage: product is intended for general intravascular use, including neurovascular and peripheral vasculatures.
Reported: November 6, 2019 Initiated: June 8, 2018 #Z-0231-2020
Product Description
Synchro2, 0.014in x 300cm STANDARD Guidewire with Hydrophilic Coating UPN Product No. M00326510, REF 2651, STERILE:EO - Product Usage: product is intended for general intravascular use, including neurovascular and peripheral vasculatures.
Reason for Recall
The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.
Details
- Recalling Firm
- Stryker Neurovascular
- Units Affected
- 2421
- Distribution
- Worldwide distributions- U.S. Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, DE, FL, GA, GU, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY. Countries of Austria, Argentina, Australia, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Lithuania, Mexico, Netherlands, Norway, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, UAE, and UK.
- Location
- Fremont, CA
Frequently Asked Questions
What product was recalled? ▼
Synchro2, 0.014in x 300cm STANDARD Guidewire with Hydrophilic Coating UPN Product No. M00326510, REF 2651, STERILE:EO - Product Usage: product is intended for general intravascular use, including neurovascular and peripheral vasculatures.. Recalled by Stryker Neurovascular. Units affected: 2421.
Why was this product recalled? ▼
The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on November 6, 2019. Severity: Moderate. Recall number: Z-0231-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11