PlainRecalls
FDA Devices Moderate Class II Terminated

SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system

Reported: November 27, 2019 Initiated: September 24, 2019 #Z-0265-2020

Product Description

SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system

Reason for Recall

SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures

Details

Units Affected
14 US
Distribution
US in the states of: GA, IN, MO, MS, NJ, OH, PA
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 14 US.
Why was this product recalled?
SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures
Which agency issued this recall?
This recall was issued by the FDA Devices on November 27, 2019. Severity: Moderate. Recall number: Z-0265-2020.