COUGAR¿ Steerable Guidewire; Models: LVCLS190J, LVCLS190S, LVCXT190J, LVCXT190S Medtronic guide wires are steerable guide wires that are used for the introduction and placement of diagnostics or interventional devices in the coronary and peripheral vasculature, and may be used to reach and cross a target lesion. Medtronic guide wires are not intended for cerebral vasculature. Medtronic steerable exchange guide wires are used to facilitate the substitution of one diagnostic or interventional d

Reported: November 27, 2013 Initiated: October 21, 2013 #Z-0266-2014

Product Description

COUGAR¿ Steerable Guidewire; Models: LVCLS190J, LVCLS190S, LVCXT190J, LVCXT190S Medtronic guide wires are steerable guide wires that are used for the introduction and placement of diagnostics or interventional devices in the coronary and peripheral vasculature, and may be used to reach and cross a target lesion. Medtronic guide wires are not intended for cerebral vasculature. Medtronic steerable exchange guide wires are used to facilitate the substitution of one diagnostic or interventional device for another

Reason for Recall

Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire.

Details

Recalling Firm
Medtronic Vascular
Units Affected
20
Distribution
Worldwide distribution: US (Nationwide) including states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV; and countries of: Albania, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guyana, Honduras, Hong Kong,Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malaysia, Mexico, Montenegro, Nepal, Netherlands, New Zealand, Nepal, Nicaragua, Norway, Oman, Poland, Portugal Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, United Kingdom, Uruguay, and Yemen.
Location
Danvers, MA

Frequently Asked Questions

What product was recalled?
COUGAR¿ Steerable Guidewire; Models: LVCLS190J, LVCLS190S, LVCXT190J, LVCXT190S Medtronic guide wires are steerable guide wires that are used for the introduction and placement of diagnostics or interventional devices in the coronary and peripheral vasculature, and may be used to reach and cross a target lesion. Medtronic guide wires are not intended for cerebral vasculature. Medtronic steerable exchange guide wires are used to facilitate the substitution of one diagnostic or interventional device for another. Recalled by Medtronic Vascular. Units affected: 20.
Why was this product recalled?
Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 27, 2013. Severity: Critical. Recall number: Z-0266-2014.