FDA Devices Moderate Class II Terminated

Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Product Usage: used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.

Reported: November 4, 2020 Initiated: September 14, 2020 #Z-0284-2021

Product Description

Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Product Usage: used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.

Reason for Recall

The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.

Details

Units Affected
8,942
Distribution
Global Distribution. US Nationwide.
Location
Andover, MA

Frequently Asked Questions

What product was recalled?
Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Product Usage: used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.. Recalled by Philips North America, LLC. Units affected: 8,942.
Why was this product recalled?
The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 4, 2020. Severity: Moderate. Recall number: Z-0284-2021.