PlainRecalls
FDA Devices Moderate Class II Terminated

4205 Journey Rolling Walker Product Usage: used to assist people in walking

Reported: November 2, 2016 Initiated: October 1, 2016 #Z-0304-2017

Product Description

4205 Journey Rolling Walker Product Usage: used to assist people in walking

Reason for Recall

Nova Medical is recalling the Journey Rolling Walker because the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out.

Details

Recalling Firm
Nova Ortho-Med Inc
Units Affected
1,380 units
Distribution
US Nationwide Distribution in the states of: CA, IN, HI, CT, UT, WA, FL, VA, WI, OH, MN, KS, MD, IL, MO, WY, MT, ME, PA, TN, MI, OR, NY, NJ, ID, NC, IA, OK, AR, VT, SC, TX
Location
Carson, CA

Frequently Asked Questions

What product was recalled?
4205 Journey Rolling Walker Product Usage: used to assist people in walking. Recalled by Nova Ortho-Med Inc. Units affected: 1,380 units.
Why was this product recalled?
Nova Medical is recalling the Journey Rolling Walker because the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 2, 2016. Severity: Moderate. Recall number: Z-0304-2017.