Alaris PCA Pump, Model 8120. The firm name on the label is CareFusion. The Alaris Syringe Pump module and Alaris PCA module are infusion pump modules for the Alaris System. The syringe pump and PCA pump delivers medication or fluids in a manner similar to current syringe pumps on the market. The modules are indicated for use on adults, pediatrics and neonates for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous, subcutaneous or ep

Reported: January 17, 2018 Initiated: November 17, 2017 #Z-0323-2018

Product Description

Alaris PCA Pump, Model 8120. The firm name on the label is CareFusion. The Alaris Syringe Pump module and Alaris PCA module are infusion pump modules for the Alaris System. The syringe pump and PCA pump delivers medication or fluids in a manner similar to current syringe pumps on the market. The modules are indicated for use on adults, pediatrics and neonates for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous, subcutaneous or epidural. The Alaris Syringe pump and PCA modules can be connected to the Alaris PC unit which is the central programming, monitoring, and power supply component for the Alaris System. The syringe pump and PCA pump uses standard, plastic piston syringes and single-use administration sets with luer-lock connectors.

Reason for Recall

The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.

Details

Recalling Firm
CareFusion 303, Inc.
Units Affected
73,611
Distribution
Worldwide Distribution - USA (nationwide) Distribution. There was also government and military distribution. Foreign distribution was made to Canada, United Arab Emirates, Bahrain, Canada, Switzerland, United Kingdom, Kuwait, Mexico, Malaysia, The Netherlands, New Zealand, Qatar, Saudi Arabia, Taiwan, and South Africa.
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
Alaris PCA Pump, Model 8120. The firm name on the label is CareFusion. The Alaris Syringe Pump module and Alaris PCA module are infusion pump modules for the Alaris System. The syringe pump and PCA pump delivers medication or fluids in a manner similar to current syringe pumps on the market. The modules are indicated for use on adults, pediatrics and neonates for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous, subcutaneous or epidural. The Alaris Syringe pump and PCA modules can be connected to the Alaris PC unit which is the central programming, monitoring, and power supply component for the Alaris System. The syringe pump and PCA pump uses standard, plastic piston syringes and single-use administration sets with luer-lock connectors.. Recalled by CareFusion 303, Inc.. Units affected: 73,611.
Why was this product recalled?
The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 17, 2018. Severity: Moderate. Recall number: Z-0323-2018.