FDA Devices Moderate Class II Terminated

RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Reported: December 8, 2021 Initiated: November 1, 2021 #Z-0328-2022

Product Description

RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Reason for Recall

When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range

Details

Recalling Firm
Randox Laboratories Ltd.
Units Affected
15 kits
Distribution
US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.
Location
Crumlin (North)

Frequently Asked Questions

What product was recalled?
RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200. Recalled by Randox Laboratories Ltd.. Units affected: 15 kits.
Why was this product recalled?
When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range
Which agency issued this recall?
This recall was issued by the FDA Devices on December 8, 2021. Severity: Moderate. Recall number: Z-0328-2022.