Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc. The iPAW ((intelligent Peri-Analytical Workstation) can act as: a software client for specimen tracking and workload statistics or a physical workstation for manual sorting and label printing.

Reported: November 7, 2018 Initiated: May 24, 2018 #Z-0353-2019

Product Description

Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc. The iPAW ((intelligent Peri-Analytical Workstation) can act as: a software client for specimen tracking and workload statistics or a physical workstation for manual sorting and label printing.

Reason for Recall

Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. Sorting Drive software version 4.2 was replaced by 4.2.1 (both versions contain the same option for manual barcode printing). When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.

Details

Recalling Firm
Beckman Coulter Inc.
Units Affected
283
Distribution
Worldwide Distribution: US (nationwide) and countries of: Algeria, Argentina, Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland, Israel, Italy ,Republic of Korea, Lebanon, Lithuania ,Malaysia, Netherlands, Poland, Portugal , Qatar, Romania, Russian Federation, Saudi Arabia ,South Africa , Spain, Sweden, Taiwan, Turkey , and United Kingdom.
Location
Brea, CA

Frequently Asked Questions

What product was recalled?
Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc. The iPAW ((intelligent Peri-Analytical Workstation) can act as: a software client for specimen tracking and workload statistics or a physical workstation for manual sorting and label printing.. Recalled by Beckman Coulter Inc.. Units affected: 283.
Why was this product recalled?
Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. Sorting Drive software version 4.2 was replaced by 4.2.1 (both versions contain the same option for manual barcode printing). When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 7, 2018. Severity: Low. Recall number: Z-0353-2019.