PlainRecalls
FDA Devices Moderate Class II Terminated

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Reported: November 20, 2019 Initiated: February 8, 2019 #Z-0360-2020

Product Description

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Reason for Recall

Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Details

Units Affected
9 systems
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.
Location
Tustin, CA

Frequently Asked Questions

What product was recalled?
Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.. Recalled by Canon Medical System, USA, INC.. Units affected: 9 systems.
Why was this product recalled?
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 20, 2019. Severity: Moderate. Recall number: Z-0360-2020.