FDA Devices Moderate Class II Ongoing

SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection,

Reported: November 5, 2025 Initiated: May 5, 2025 #Z-0372-2026

Product Description

SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910

Reason for Recall

All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.

Details

Units Affected
4,272,858
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.
Location
Austell, GA

Frequently Asked Questions

What product was recalled?
SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910. Recalled by ASP Global, LLC. dba Anatomy Supply Partners, LLC.. Units affected: 4,272,858.
Why was this product recalled?
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Which agency issued this recall?