SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection,
Product Description
SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910
Reason for Recall
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Details
- Recalling Firm
- ASP Global, LLC. dba Anatomy Supply Partners, LLC.
- Units Affected
- 4,272,858
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.
- Location
- Austell, GA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11