EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V
Reported: December 22, 2021 Initiated: November 2, 2021 #Z-0379-2022
Product Description
EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V
Reason for Recall
Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.
Details
- Recalling Firm
- Olympus Corporation of the Americas
- Units Affected
- 14,447 units
- Distribution
- US Nationwide distribution.
- Location
- Center Valley, PA
Frequently Asked Questions
What product was recalled? ▼
EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V. Recalled by Olympus Corporation of the Americas. Units affected: 14,447 units.
Why was this product recalled? ▼
Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on December 22, 2021. Severity: Moderate. Recall number: Z-0379-2022.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11