PlainRecalls
FDA Devices Moderate Class II Terminated

. 00789502801 VerSys Femoral Head Provisional 28mm -3.5 00789502802 VerSys Femoral Head Provisional 28mm +0 00789502803 VerSys Femoral Head Provisional 28mm +3.5 00789502804 VerSys Femoral Head Provisional 28mm +7 00789502805 VerSys Femoral Head Provisional 28mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty

Reported: November 28, 2012 Initiated: October 10, 2010 #Z-0391-2013

Product Description

. 00789502801 VerSys Femoral Head Provisional 28mm -3.5 00789502802 VerSys Femoral Head Provisional 28mm +0 00789502803 VerSys Femoral Head Provisional 28mm +3.5 00789502804 VerSys Femoral Head Provisional 28mm +7 00789502805 VerSys Femoral Head Provisional 28mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty

Reason for Recall

Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat

Details

Recalling Firm
Zimmer, Inc.
Units Affected
128,946 all lots, all sizes
Distribution
Worldwide distribution: USA (nationwide) and countries of: Asia, Barbados, Canada, Europe, Mexico, Puerto Rico, San Juan, and South America.
Location
Warsaw, IN

Frequently Asked Questions

What product was recalled?
. 00789502801 VerSys Femoral Head Provisional 28mm -3.5 00789502802 VerSys Femoral Head Provisional 28mm +0 00789502803 VerSys Femoral Head Provisional 28mm +3.5 00789502804 VerSys Femoral Head Provisional 28mm +7 00789502805 VerSys Femoral Head Provisional 28mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty. Recalled by Zimmer, Inc.. Units affected: 128,946 all lots, all sizes.
Why was this product recalled?
Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat
Which agency issued this recall?
This recall was issued by the FDA Devices on November 28, 2012. Severity: Moderate. Recall number: Z-0391-2013.