ORTHOFIX Catalogue Number: ref: 99-93502, LOWER LIMB DIAPHYSEAL STERILE KIT, RX ONLY, UDI: (01)18033509859915
Reported: November 20, 2019 Initiated: September 27, 2018 #Z-0397-2020
Product Description
ORTHOFIX Catalogue Number: ref: 99-93502, LOWER LIMB DIAPHYSEAL STERILE KIT, RX ONLY, UDI: (01)18033509859915
Reason for Recall
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Details
- Recalling Firm
- Orthofix Srl
- Units Affected
- N/A
- Distribution
- US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
- Location
- Veronaverona, N/A
Frequently Asked Questions
What product was recalled? ▼
ORTHOFIX Catalogue Number: ref: 99-93502, LOWER LIMB DIAPHYSEAL STERILE KIT, RX ONLY, UDI: (01)18033509859915. Recalled by Orthofix Srl. Units affected: N/A.
Why was this product recalled? ▼
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on November 20, 2019. Severity: Moderate. Recall number: Z-0397-2020.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11