Brilliance iCT - Model 728306 Computed Tomography X-ray systems

Reported: January 31, 2018 Initiated: July 11, 2017 #Z-0403-2018

Product Description

Brilliance iCT - Model 728306 Computed Tomography X-ray systems

Reason for Recall

Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.

Details

Units Affected
559 total
Distribution
Worldwide distribution, including US nationwide.
Location
Cleveland, OH

Frequently Asked Questions

What product was recalled?
Brilliance iCT - Model 728306 Computed Tomography X-ray systems. Recalled by Philips Medical Systems (Cleveland) Inc. Units affected: 559 total.
Why was this product recalled?
Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.