PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
Reported: December 6, 2023 Initiated: October 16, 2023 #Z-0419-2024
Product Description
PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
Reason for Recall
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
Details
- Recalling Firm
- Aizu Olympus Co., Ltd.
- Units Affected
- 24 incorrectly repaired units
- Distribution
- Domestic distribution to FL, KY, MA, NJ, NY, and PA
- Location
- Aizuwakamatsu
Frequently Asked Questions
What product was recalled? ▼
PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE. Recalled by Aizu Olympus Co., Ltd.. Units affected: 24 incorrectly repaired units.
Why was this product recalled? ▼
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on December 6, 2023. Severity: Moderate. Recall number: Z-0419-2024.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11