Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635
Reported: January 5, 2022 Initiated: December 1, 2021 #Z-0440-2022
Product Description
Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635
Reason for Recall
¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- US: 46 units OUS: 794 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, CO, MA, MD, MS, NC, NE, NH, OH, VA, WI and the countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Czech Republic, Denmark, Ecuador, France, Germany, Greece, India, Iran, Israel, Italy, Japan, Mexico, Morocco, Oman, P.R. China, Pakistan, Philippines, Portugal, Slovakia, South Africa, Spain, Switzerland, Turkey, United Kingdom, Uruguay.
- Location
- Tarrytown, NY
Frequently Asked Questions
What product was recalled? ▼
Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: US: 46 units OUS: 794 units.
Why was this product recalled? ▼
¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on January 5, 2022. Severity: Moderate. Recall number: Z-0440-2022.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11