BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251
Product Description
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251
Reason for Recall
Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
Details
- Recalling Firm
- BioFire Diagnostics, LLC
- Units Affected
- 32,910
- Distribution
- Worldwide Distribution: US (nationwide): DE, OH, NY, IL, KY, FL, GA, KS, CA, WI, NC,, AK, PA, MN, ND, IN, AR, CO, UT, TX, AL, NJ, MD, SD, LA, OR, MA, ID, NE, MT, VA, IA, MO, WV, TN, MI, WY, SC, AZ, WA, NH, OK, HI, MS, DC, VT, CT, NM, RI, ME, NV; and OUS (foreign) to countries of: Algeria, Angola, Armenia, Bahrain, Bolivia, Bosnia- Herzegovina, Brunei, Bulgaria, Costa Rica, Croatia, Cyprus, Dominican Republic, Dutch Antilles, Ecuador, El Salvador, Estonia, Gabon, Georgia, Guam, Guatemala, Honduras, Indonesia, Ireland, Iceland, Israel, Ivory Coast, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macao, Macedonia, Malaysia, Mali, Mongolia, Morocco, New Zealand, Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Qatar, Romania, Saudi Arabia, Serbia , Slovenia, Tunisia, Uganda, United Arab Emirate, Uruguay, Venezuela, Vietnam, Zimbabwe, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Spain, France, United Kingdom, Greece, India, Italia, Japan, Korea, Netherlands, Poland, Portugal, Sweden, Singapore, Thailand, Turkey, Taiwan, Switzerland, Austria, Denmark, Finland, Germany, Belgium, Hungary, South Africa, Philippines, Norway, Mexico
- Location
- Salt Lake City, UT
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11