PlainRecalls
FDA Devices Moderate Class II Terminated

Dimension Vista(R) IRON Flex(R) reagent cartridge (K3085) The IRON method is an in vitro diagnostic test for the quantitative determination of iron in human serum and plasma on the Dimension Vista(R) System. Measurements of iron are used in the diagnosis and treatment of diseases such as iron deficiency anemia and other disorders of iron metabolism.

Reported: December 12, 2012 Initiated: October 8, 2012 #Z-0452-2013

Product Description

Dimension Vista(R) IRON Flex(R) reagent cartridge (K3085) The IRON method is an in vitro diagnostic test for the quantitative determination of iron in human serum and plasma on the Dimension Vista(R) System. Measurements of iron are used in the diagnosis and treatment of diseases such as iron deficiency anemia and other disorders of iron metabolism.

Reason for Recall

Siemens Healthcare Diagnostics has confirmed the potential for falsely elevated results with IRON Flex(R) reagent cartridges when tests are processed from a well set contaminated with iron from environmental sources. All lots of IRON Flex(R) reagent cartridges have the potential for this issue. If a well set is contaminated with trace amounts of iron, the entire well set is impacted, and the mag

Details

Units Affected
19,310
Distribution
Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, Washington DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and Internationally to Canada, AU, NZ, JP, DE, SA, and MY
Location
Newark, DE

Frequently Asked Questions

What product was recalled?
Dimension Vista(R) IRON Flex(R) reagent cartridge (K3085) The IRON method is an in vitro diagnostic test for the quantitative determination of iron in human serum and plasma on the Dimension Vista(R) System. Measurements of iron are used in the diagnosis and treatment of diseases such as iron deficiency anemia and other disorders of iron metabolism.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 19,310.
Why was this product recalled?
Siemens Healthcare Diagnostics has confirmed the potential for falsely elevated results with IRON Flex(R) reagent cartridges when tests are processed from a well set contaminated with iron from environmental sources. All lots of IRON Flex(R) reagent cartridges have the potential for this issue. If a well set is contaminated with trace amounts of iron, the entire well set is impacted, and the mag
Which agency issued this recall?
This recall was issued by the FDA Devices on December 12, 2012. Severity: Moderate. Recall number: Z-0452-2013.