Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System

Reported: December 12, 2012 Initiated: June 20, 2012 #Z-0465-2013

Product Description

Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System

Reason for Recall

Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could

Details

Units Affected
15
Distribution
Nationwide distribution: USA including states of: CA, FL, GA, MA, MN, MO, NJ, NY, OH, and WI.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 15.
Why was this product recalled?
Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could
Which agency issued this recall?
This recall was issued by the FDA Devices on December 12, 2012. Severity: Moderate. Recall number: Z-0465-2013.