PlainRecalls
FDA Devices Moderate Class II Terminated

GLOBAL UNITE STD STEM SZ 12 Product Code: 110012100

Reported: January 19, 2022 Initiated: December 7, 2021 #Z-0488-2022

Product Description

GLOBAL UNITE STD STEM SZ 12 Product Code: 110012100

Reason for Recall

Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage

Details

Recalling Firm
DePuy Orthopaedics, Inc.
Units Affected
870 units
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Netherlands, New Zealand, Norway, Poland, South Africa, Sweden, Switzerland, Thailand, United Kingdom.
Location
Warsaw, IN

Frequently Asked Questions

What product was recalled?
GLOBAL UNITE STD STEM SZ 12 Product Code: 110012100. Recalled by DePuy Orthopaedics, Inc.. Units affected: 870 units.
Why was this product recalled?
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
Which agency issued this recall?
This recall was issued by the FDA Devices on January 19, 2022. Severity: Moderate. Recall number: Z-0488-2022.