AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reported: February 14, 2018 Initiated: August 24, 2017 #Z-0508-2018

Product Description

AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason for Recall

Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.

Details

Units Affected
265 units
Distribution
US Nationwide Distribution
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 265 units.
Why was this product recalled?
Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 14, 2018. Severity: Moderate. Recall number: Z-0508-2018.