Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system

Reported: February 14, 2018 Initiated: November 1, 2017 #Z-0522-2018

Product Description

Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system

Reason for Recall

One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware.

Details

Units Affected
2
Distribution
One domestic and 1 foreign consignee ( Spain)
Location
Cleveland, OH

Frequently Asked Questions

What product was recalled?
Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system. Recalled by Philips Medical Systems (Cleveland) Inc. Units affected: 2.
Why was this product recalled?
One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 14, 2018. Severity: Moderate. Recall number: Z-0522-2018.