smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP

Reported: December 4, 2024 Initiated: October 28, 2024 #Z-0528-2025

Product Description

smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP

Reason for Recall

Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.

Details

Recalling Firm
Smiths Medical ASD, Inc.
Units Affected
10,223 units
Distribution
Worldwide distribution.
Location
Minneapolis, MN

Frequently Asked Questions

What product was recalled?
smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP. Recalled by Smiths Medical ASD, Inc.. Units affected: 10,223 units.
Why was this product recalled?
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 4, 2024. Severity: Moderate. Recall number: Z-0528-2025.