PlainRecalls
FDA Devices Moderate Class II Terminated

LT DISC ULNA - HA/PC; RT DISC ULNA - HA/PC; LT DISC HMRL- HA/PC; RT DISC HMRL- HA/PC; DISC XS ULNA POR; DISC SEG ULNA POR; and Custom devices Item Nos. 114754 114755 114744 114745 114704 114705 114756 114757 114714 114715 114746 114747 114706 114707 114758 114759 114716 114717 114748 114749 114708 114709 114718 114719 CP561429 CP561430 CP561431 CP561432 CP561442 CP561449 CP561488 CP561489 CP561709 CP561710 CP561714 CP561715 PM555203 Product Usage:

Reported: December 4, 2019 Initiated: August 21, 2019 #Z-0532-2020

Product Description

LT DISC ULNA - HA/PC; RT DISC ULNA - HA/PC; LT DISC HMRL- HA/PC; RT DISC HMRL- HA/PC; DISC XS ULNA POR; DISC SEG ULNA POR; and Custom devices Item Nos. 114754 114755 114744 114745 114704 114705 114756 114757 114714 114715 114746 114747 114706 114707 114758 114759 114716 114717 114748 114749 114708 114709 114718 114719 CP561429 CP561430 CP561431 CP561432 CP561442 CP561449 CP561488 CP561489 CP561709 CP561710 CP561714 CP561715 PM555203 Product Usage: Total Elbow Replacement

Reason for Recall

Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.

Details

Recalling Firm
Zimmer Biomet, Inc.
Units Affected
217240
Distribution
US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
Location
Warsaw, IN

Frequently Asked Questions

What product was recalled?
LT DISC ULNA - HA/PC; RT DISC ULNA - HA/PC; LT DISC HMRL- HA/PC; RT DISC HMRL- HA/PC; DISC XS ULNA POR; DISC SEG ULNA POR; and Custom devices Item Nos. 114754 114755 114744 114745 114704 114705 114756 114757 114714 114715 114746 114747 114706 114707 114758 114759 114716 114717 114748 114749 114708 114709 114718 114719 CP561429 CP561430 CP561431 CP561432 CP561442 CP561449 CP561488 CP561489 CP561709 CP561710 CP561714 CP561715 PM555203 Product Usage: Total Elbow Replacement. Recalled by Zimmer Biomet, Inc.. Units affected: 217240.
Why was this product recalled?
Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 4, 2019. Severity: Moderate. Recall number: Z-0532-2020.