Covidien Vistec" X-Ray Detectable Sponges, 7317 Vistec" X-Ray Detectable Sponges Sponge. product Usage: Vistec X-Ray Detectable Sponges are medical absorbent devices intended to be used internally within the body cavity to control bleeding or for applying medication. These sponges contain unique dual-colored elements that combines blue radiopaque strand with white strand, which is visible when sponge is saturated with blood.

Reported: February 14, 2018 Initiated: July 17, 2017 #Z-0534-2018

Product Description

Covidien Vistec" X-Ray Detectable Sponges, 7317 Vistec" X-Ray Detectable Sponges Sponge. product Usage: Vistec X-Ray Detectable Sponges are medical absorbent devices intended to be used internally within the body cavity to control bleeding or for applying medication. These sponges contain unique dual-colored elements that combines blue radiopaque strand with white strand, which is visible when sponge is saturated with blood.

Reason for Recall

Medtronic is voluntarily recalling two production lots of its Covidien Vistec" x-ray detectable sponges de to possible contamination with human blood caused by handling of the product during the manufacturing process and prior to final packaging and Ethylene Oxide sterilization.

Details

Recalling Firm
Covidien LP
Units Affected
33,280
Distribution
U.S., Gov Accts, and foreign countries: PA, PR, CO.
Location
Augusta, GA

Frequently Asked Questions

What product was recalled?
Covidien Vistec" X-Ray Detectable Sponges, 7317 Vistec" X-Ray Detectable Sponges Sponge. product Usage: Vistec X-Ray Detectable Sponges are medical absorbent devices intended to be used internally within the body cavity to control bleeding or for applying medication. These sponges contain unique dual-colored elements that combines blue radiopaque strand with white strand, which is visible when sponge is saturated with blood.. Recalled by Covidien LP. Units affected: 33,280.
Why was this product recalled?
Medtronic is voluntarily recalling two production lots of its Covidien Vistec" x-ray detectable sponges de to possible contamination with human blood caused by handling of the product during the manufacturing process and prior to final packaging and Ethylene Oxide sterilization.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 14, 2018. Severity: Moderate. Recall number: Z-0534-2018.