Medfusion¿ 4000 Syringe Infusion Pumps Medfusion¿ 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.

Reported: December 25, 2013 Initiated: November 22, 2013 #Z-0540-2014

Product Description

Medfusion¿ 4000 Syringe Infusion Pumps Medfusion¿ 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.

Reason for Recall

Smiths Medical is conducting a recall involving all Medfusion¿ 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po

Details

Recalling Firm
Smiths Medical ASD, Inc.
Units Affected
9560 (8601 US, 959 OUS)
Distribution
Worldwide Distribution - USA (nationwide ) including the states of : AZ, NY, TN, MS, AR, WV, VA, VT, PA, DE, DC, OH, CO, NH, MI, CA, FL, CT, WA, MD, NM, RI, MA, TX, NC, IL, KY, WI and MI., and the countries of CANADA, SAUDI ARABIA and AUSTRALIA.
Location
Saint Paul, MN

Frequently Asked Questions

What product was recalled?
Medfusion¿ 4000 Syringe Infusion Pumps Medfusion¿ 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.. Recalled by Smiths Medical ASD, Inc.. Units affected: 9560 (8601 US, 959 OUS).
Why was this product recalled?
Smiths Medical is conducting a recall involving all Medfusion¿ 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po
Which agency issued this recall?
This recall was issued by the FDA Devices on December 25, 2013. Severity: Moderate. Recall number: Z-0540-2014.