Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Reported: December 4, 2024 Initiated: October 9, 2024 #Z-0554-2025
Product Description
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Reason for Recall
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
Details
- Recalling Firm
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Units Affected
- 48
- Distribution
- Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
- Location
- Kyoto
Frequently Asked Questions
What product was recalled? ▼
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL. Recalled by Shimadzu Corp. Analytical & Measuring Instruments Div.. Units affected: 48.
Why was this product recalled? ▼
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on December 4, 2024. Severity: Moderate. Recall number: Z-0554-2025.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11